You could save on the cost of some of the Protocol 1 calibration gases and the
shipping of same by following Test Method 205 (linked here).
This helpful procedure describes the use of calibration gas dilution devices to produce
numerous calibration gases for just such instrument checks without lugging
all those cylinders to the test site.
Purified ambient air is still allowed for Method 3A.
Mention of it was an inadvertent omission in the final
revisions. This provision will be added back in the
correction notice. The purified air should be free
of any interfering components.
Equation 7E-9 in the final rule appears to be correct;
at least, it's essentially the same equation taken from old Method 20.
Still have questions on this method? Contact the EMC
expert Foston Curtis at
curtis.foston@epa.gov .