Method 5 - Particulate Matter (PM)

Text of Test Method 5 - Particulate Matter (PM) PDF file
Technical Information Document (TID) 001 - Derivation of Delta H PDF file
Technical Information Document (TID) 019 - Clarification of Method 5, 201A, and 6 Requirements PDF file
Method 5 (Method 0010) Sampling Train txt file
Maintenance, Calibration & Operation of Isokinetic Source Sampling Equipment - APTD-0576 PDF file
Construction Details of Isokinetic Source Sampling Equipment - APTD-0581 PDF file
Alternative Method 5 Post-Test Calibration PDF file
Particulate (PM Screening) Method (CTM-002) PDF file
Particulate (PM Screening) Method Figures PDF file
Particulate (PM Modified High Vol.) Method (CTM-003) PDF file
Electronic Reporting Tool



FREQUENTLY ASKED QUESTIONS (FAQS) for Method 5



Does Section 4.2 of Method 5 of 40 CFR Part 60 require a 200 ml sample for each of the three test runs collected during a performance test?

We Interpret this section of the method to allow the determination of a blank value for the acetone rinse liquid on a lot basis rather than requiring a 200 ml sample for each test run. One or more bottles (eg., gallon bottles) usually constitutes a lot and is identified by a lot number. To determine and document the blank value for a test and a lot of acetone, the tester must:
1. Collect a 200 ml sample of acetone for each lot of acetone used during the test from the acetone wash bottle used for the sample recovery and analyze the sample as per the requirements in sections 4.2 and 4.3 of Method 5;
2. Use only one acetone wash bottle per lot per test;
3. Record the lot number of the acetone used during the sample recovery for each test run on the applicable run data sheet;
4. Avoid mixing lots of acetone in the sample recovery for any one test run by using the acetone from only one lot in cleaning the probe and filter holder following each test run; and
5. Apply the acetone blank value for the applicable acetone lot in the calculations for each test run. If acetone from more than one lot is used during a test (e.g., lot A for run 1 and lot B for runs 2 and 3), apply the appropriate blank value for each test run.

We have been getting complaints from the State inspectors about the humidity in our filter weighing room not meeting Method 5 Criteria. I have read those criteria and can not find anything specific about the weighing room conditions. Are there any specs?

There are criteria in the method, Section 4.1.1, that may not be all that apparent but are also not too hard to meet. Basically, the filter weighing area must have a relative humidity of less than 50 percent. Most air conditioned rooms meet this easily. More importantly, the method specifies that the filter not be exposed to any such room conditions more than 2 minutes for the weighing process. My interpretation is that this latter criterion applies whether the filter is dried in a dessicator or is oven-dried.

Some of the places we test have such low particulate emission concentrations that, to collect 50 mg of sample, we would have to test into next week. What is the lower limit on a Method 5 catch for a test to be accepted? Can we collect less than 10 mg of catch and still show compliance?

This is a good question and frequently asked - one for which EPA has not developed a definitive answer. I know that EPA uses a minimum catch of about 50 mg (front half) when calculating the minimum sample volume required to show compliance with new regulations. That is, EPA calculates how many cubic feet of sample are necessary to collect 50 mg on the probe and filter for a source emission concentration at the level of the standard. Then, even though the tester may collect much less than 50 mg during the compliance test, the enforcement agency can be pretty sure that the source is in compliance using a three-run average.
On the other hand, there are data indicating that very acceptable precision and repeatability (within about 0.1 mg) can be achieved for catches around 10 mg. In some cases, compliance status could be established with sample catches at these levels if the number of test runs were increased to six or nine. Collection of even this small amount of sample can become a problem when process operation is very short, dictating that the test run also be very short. Special testing schemes must be developed for these cases.
Of course, there are good practices testers should follow in collecting and analyzing particulate samples. These include safeguards such as using low residue solvents, ensuring clean wash bottles and brushes, using low tare plastic or Teflon beakers and petri dishes, and careful handling in the lab. Repeating sample brushing, rinsing, and collection procedures three and four times in the interest of thoroughness can only improve the results.

What is the importance of the Method 5, Section 7.1.1 specification that the filter material be “unreactive” to SO2 or SO3?

Method 5, Section 7.1.1, states, “In sources containing SO2 or SO3, the filter material must be of a type that is unreactive to SO2 or SO3. Reference 10 in Section 17.0 may be used to select the appropriate filter.” The reaction of gaseous compounds such as SO2 and SO3 with filter media can result in large positive errors in mass on the filters. As the amount of particulate matter sampled using Method 5 decreases, it becomes increasingly important to ensure filter materials are unreactive to the constituents of the sample matrix.
The language in Section 7.1.1 was added to Method 5 on October 7, 1980 (45 FR 66742). Comments were received during the promulgation process for the NSPS for Glass Manufacturing Plants describing “inflated particulate emission values” when sampling occurred in the presence of SOx. Based on these comments and an EPA report released in June 1977, “EPA-600/7-77-06 - Inertial Cascade Impactor Substrate Media for Flue Gas Sampling, linked here, Method 5 was updated to require the use of “non-reactive filters” at sources containing SO2 or SO3, and to suggest that the above mentioned report be used to select the appropriate filter material.
The EPA report EPA-600/7-77-06 (cited above) states, “Substrate mass gains have been found to be a source of very large errors when sampling industrial flue gases with inertial impactors in which glass fiber material is used for substrates. Ideally, the only mass change in a substrate should be that resulting from collection of particulate matter from the flue gas. However, if gas-phase reactions take place, involving components of the glass fiber substrate, then it is possible for substantial mass gains to occur, unrelated to the particle size distribution.”
Additionally, a bias has also been discovered in the presence of HCl. Two reports, “Laboratory and Field Evaluation of a Methodology for Determination of Hydrogen Chloride Emissions from Municipal and Hazardous Waste Incinerators by Scott Steinsberger and John Margeson, linked here, and “Stack Sampling Methods for Halogens and Halogen Acids by Larry D. Johnson, linked here, discuss this issue and recommend the use of “non-reactive filters” or quartz filters in the presence of HCl, also.
In order to avoid the possibility of biased results, some testers have decided to use only quartz filters when performing Method 5.
More information about filter reactivity and the selection of non-reactive filters can be found in two documents published in the United Kingdom, “Study into the loss of material from filters used for collecting particulate matter during stack emissions monitoring,” linked here, Exit EPA disclaimer
and “Method Implementation Document for EN 13284, BS EN 13284-1:2002, Stationary source emissions - Determination of low range mass concentration of dust - Part 1: Manual gravimetric method,” linked here. Exit EPA disclaimer

Do you have questions about a Test Method, Performance Specification or Source Category? Find the EMC contact for your question by clicking on the "EMC Contacts" link at the menu on the left of your screen, or use the direct link here.